How Do You Solve These Faults Of The Hematology Analyzer?

What Common Problems Will Occur During The Use Of Blood Analyzers?

Hematology analyzer, also known as blood cell analyzer, blood cell counter, etc., is widely used in clinical practice in hospitals, and it is also very easy to fail during use. Toady CN MEDITECH will introduce the common faults and maintenance methods of hematology analyzers.Hope we can help you to solve these troubles!

hematology analyzer

1.Background error when counting. When the automatic detection blank value exceeds the specified value, an alarm will be prompted, and a background error occurs at this time. At this time, you can first check whether the hemolytic agent or diluent has air bubbles and whether it is contaminated. If so, you need to replace the reagent in time; in addition, you can check whether there is electromagnetic wave interference in the equipment to ensure that the equipment is well grounded. This fault can also occur if the pores of the detection chamber are contaminated. At this time, the maintenance and cleaning program can be started for cleaning. If the above reasons are ruled out, you can check whether the rotary valve is contaminated, so as to deal with it accordingly.

2.RBC/WBC/PLT count error. You can first check whether the fault is caused by sample agglutination, and then check whether it receives noise, electromagnetic wave interference, or poor grounding. Abnormal emptying of the insulating chamber below the blood cell detection chamber can also cause this fault. In addition, check whether the pores of the detection chamber or the detector are polluted or damaged, whether the pipeline between the cell counting quantitative diaphragm pump and the detector is blocked, whether the circuit board is faulty, etc., should be eliminated one by one.

3.WBC counting time is abnormal. When the counting time is too long, it may be because the counting hole is polluted, or it may be because the blood contains fibrin crystals and tiny aggregates that cause the hole to be blocked. At the same time, you can check whether there is a problem with the air injector due to air leakage from the solenoid valve of the WBC pipeline. And cause the negative pressure to be too low. When the counting time is too short, it may be due to the failure of the liquid circuit. You should check whether the liquid circuit tube is squeezed, whether the counting pool is firm, and whether the sample outlet of the dilution pool is blocked.

What Aspects Should Be Paid Attention To When We Use Hematology Analyzer?

blood analyzer

1.Blood sample: Since venous blood is less affected by external factors, the composition is stable, and the accuracy and repeatability of the results are good, it is recommended to use venous blood for testing except for infants. At the same time, because the composition of the first drop of blood is unstable, it should be Tested with a second drop of blood. When testing peripheral blood, be careful not to over-squeeze the area to avoid mixing with interstitial fluid, thereby activating the coagulation system and causing errors in the test results.

2.Anticoagulation: If the citrate anticoagulant is used for too long, crystallization will easily occur, and the cell morphology will easily change, which will affect the accuracy of the counting results; oxalate will easily cause platelet aggregation, which will change the morphology of white blood cells and affect the counting Results and classification; excessive heparin can easily cause leukocyte aggregation and thrombocytopenia; EDTA-2Na is less soluble than EDTA-2K, and the possibility of platelet aggregation is greater. Therefore, the document published by the International Committee for Hematology Standards (ICSH) in 1993 suggested using EDTA-2K as an anticoagulant for blood analyzers.

3.Dilution: Before use, the diluter and the suction tube should be calibrated, and the blood outside the suction tube should be completely wiped clean after use. It should be measured as soon as possible after blood dilution, otherwise it will easily cause “dilution hemolysis”.

4.Mixing: It is very important to mix before testing. If you use a non-rotating mixer, you should invert and mix at least 8 times.

5.Reagents: Blood analyzers have very strict requirements on reagents when used, requiring strict osmotic pressure standards, stable conductivity, high standards of purity, and no corrosion to instrument pipelines and valve circuits. Therefore, the hemolytic agent, diluent and cleaning agent are best to use the original supporting products.

6.White blood cell classification: At present, the most advanced blood analyzers in the world are just a screening method and cannot completely replace manual microscopy classification. We must resolutely correct the wrong idea of abandoning microscopy when using blood analyzers.

7.Blood storage: After the blood collection is completed, it needs to be sealed and stored at a suitable temperature and should not exceed 6 hours.

8.Quality control: In the process of clinical testing, a strict quality control system must be established to ensure the accuracy of the results.

In short, blood analyzers are one of the most important instruments in clinical testing. During use, the operating procedures and precautions must be strictly followed to ensure the accuracy and reliability of blood analysis results.

Why Choose Us?

CNMEDITECH is dedicated to the long-term research of the hematology analyzer market. Our mission is “People oriented and win-win strategy,Matching the real needs of the region with a focus on human health,To be the world’s first-class medical field solution expert”. We have been manufacturing high-quality medical device products for more than ten years.

We have built our reputation on delivering quality healthcare solutions on time and on budget. All our products comply with international health and safety regulations and all products come with a warranty.

Are you still worrying about your customer’s product needs? Are you still angry that the product is expensive?Our hematology analyzers have various models and support personalized product customization.

Our company has many styles to choose from. In addition, we have high-quality pre-sales consulting guidance and professional after-sales service, all to meet your needs.

Whether it is a cost-effective or high-end product, there will always be something suitable for you. If you have any needs for products, you can ask us, our factory will meet your needs as soon as possible, and we will make every effort to provide you with solutions.Feel free to send us your inquiries.

hematology analyzer

Leave a Reply

Your email address will not be published. Required fields are marked *

don't hesitate to get a quote

Get a quote now

We will contact you within 1 working day.